Some Useful Information Concerning the results:
The Conversion factors for the primordial nuclides are given by: 1 Bq 238U/kg = 81 ppb U (81 x 10-9 gU/g) The 238U decay chain gammas used are: 234Th: 63.29 and 92.59 keV
226Ra: 186.1 keV
214Pb: 295.21 and 351.92 keV
214Bi: 609.31, 1120.29, 1764.49 and 2204.21 keV
The relationships are valid for any daughters in the 238U, 235U or 232Th chain only if the chain is in equilibrium. 1 Bq 232Th/kg = 246 ppb Th (246 x 10-9 gTh/g) The 232Th decay chain gammas used are: 212Pb: 238.63 and 300.09 keV
208Tl: 277.371, 583.19, 860.557 and 2614.53 keV,
228Ac: 209.253, 338.320, 463,004, 911.21, 964.766 and 968.971 keV
1 Bq 40K/kg = 32300 ppb K (32300 x 10-6 gK/g) The 40K decay chain gamma used is: 40K: 1460.83 keV
1 Bq 235U/kg = 1.76 ppm U (1.76 x 10-6 gU/g) The 235U decay chain gammas used are: 235U: 143.76, 163.33 and 205.31 keV
Background runs for the Lively Detector
If a measurement in the signal region is below the sideband regions then the 90% confidence limit is calculated.
Sample Description Manufacturer Mass (g) Live Time (days) Run Numbers Counting Dates
(if applicable)























Background 1

171.327
Mar 29, 2019 Results:
238U from 226Ra

238U from 234Th

235U

232Th

40K

137Cs

60Co
Comments
Completely Empty Detector



Jan 20, 2020 (mBq) 1.201 +- 0.053 2.58 +- 0.40 0.110 +- 0.015 1.074 +- 0.058 4.97 +- 0.45 0.024 +- 0.021 0.118 +- 0.018





































Results: 210Pb:
7Be:

54Mn

228Ac:

210Po:














(mBq) 289.25 +- 31.53 <0.26

0.024 +- 0.021 1.305 +- 0.099 <2629.00









































Background 2

55.461 200324
200430
Mar 24, 2020 Results:
238U from 226Ra

238U from 234Th

235U

232Th

40K

137Cs

60Co
Comments
Completely Empty Detector



May 19, 2020 (mBq) 1.243 +- 0.083 3.02 +- 0.86 0.161 +- 0.029 1.38 +- 0.10 8.35 +- 0.94 <0.043

0.111 +- 0.031





































Results: 210Pb:
7Be:

54Mn

228Ac:

210Po:














(mBq) 554.52 +- 80.36 0.74 +- 0.30 0.020 +- 0.035 2.22 +- 0.18 <4463.00









































Combined Background Combined Backgrounds of runs CW1
171.327

Results:
238U from 226Ra

238U from 234Th

235U

232Th

40K

137Cs

60Co
Comments






(mBq) 1.201 +- 0.053 2.58 +- 0.40 0.110 +- 0.015 1.074 +- 0.058 4.97 +- 0.45 0.024 +- 0.021 0.118 +- 0.018
Completely Empty Detector



































Results: 210Pb:
7Be:

54Mn

228Ac:

210Po:














(mBq) 289.25 +- 31.53 <0.26

0.024 +- 0.021 1.305 +- 0.099 <2629.00









































Completed Sample Measurements for the Lively Detector
DAMIC Measurements:




























Sample Description
Mass (g) Live Time (days) Run Numbers Counting Dates
(if applicable)























DAMIC L01 3 pieces of OHFC copper 372.43 g 20.554 221219 Dec 19, 2022 Results:
238U from 226Ra

238U from 234Th

235U

232Th

40K

137Cs

60Co
Comments
OHFC Copper



Jan 9, 2023 (mBq/kg) <0.055

<5.55

<0.076

<0.27

<3.61

0.11 +- 0.11 <0.12








(ppm / ppb / ppt) <4.46 ppt

<449.55 ppt

<133.76 ppt
. <66.42 ppt

<116.6 ppb
















Results: 210Pb:
7Be:

54Mn

228Ac:
57Co:
58Co:












(mBq/kg) <205.40

<1.43

<0.19

0.43 +- 0.41 <0.17

0.37 +- 0.13











(ppm / ppb / ppt) <16.64 ppb







106.79 ppt +- 100.98 ppt











In Progress Sample Measurements for the Lively Detector
Runs in Progress:




























Sample Description
Mass (g) Live Time (days) Run Numbers Counting Dates
(if applicable)























Next Sample




Results:
238U from 226Ra

238U from 234Th

235U

232Th

40K

137Cs

60Co
Comments






(mBq/kg)



























(ppm / ppb / ppt)







.




















Results: 210Pb:
7Be:

54Mn

228Ac:
210Po:














(mBq/kg)





























(ppm / ppb / ppt)